A red status indicator almost always means the AED has failed its automatic self-test and should be treated as out of service until the fault is cleared. The important exception is that red does not mean the same thing on every device. On a HeartSine samaritan PAD a brief red flash during the Sunday self-test is normal. On Philips HeartStart units a fault shows as no light at all, never red. Before you act, confirm what your specific model is actually telling you.

This guide decodes the status indicator for each major AED brand, explains the three cases where red is not a fault, and sets out what to check and in what order. It is about reading the indicator correctly. If you have already confirmed a genuine self-test failure and need the brand-by-brand fix steps, those are in the guide on what to do when an AED self-test fails.

2Major brands where a fault shows as no light, not red
7 daysDefault ZOLL AED Plus self-test interval, not daily
1Brand where a weekly red flash is expected behaviour

What the status indicator is actually reporting

Every modern AED runs an unattended self-test on a fixed schedule. The indicator on the outside of the case is a summary of the most recent result, nothing more. It is not a live reading, and it is not a measurement taken at the moment you looked at it.

That distinction matters more than it sounds. A device that failed its test four days ago has been showing a fault for four days, and a device that passed its test yesterday can have developed a problem since. The indicator tells you the outcome of the last scheduled check, which is why the self-test interval in the table below is as important as the colour.

Status indicator meanings by brand

The ready state, the fault state, the sound and the test schedule all differ by manufacturer. The table below is drawn from each manufacturer’s own operating instructions.

Device Ready Needs attention Sound Self-test schedule
ZOLL AED Plus Green check mark in the status window Red X in the status window Audible alert with the red X Every 7 days by default, configurable from 1 to 7 days. Software 5.32 and above also runs a monthly test
Philips HeartStart OnSite (HS1) Blinking green ready light Ready light off, not red Single chirps mean attention needed, press the blue i-button. Triple chirps mean a serious fault Daily, plus a self-test whenever a battery is inserted
Philips HeartStart FRx Blinking green ready light Ready light off, not red Single chirps with a flashing i-button point to a self-test error, pads problem, a left-in Infant/Child Key or low battery. Triple chirps mean a serious fault Daily, plus a battery insertion test
Stryker LIFEPAK CR2 Readiness indicator flashes every 6 seconds Indicator stops flashing Alert tone every 15 minutes, which can be disabled in setup Automatic, with status also reported through LIFELINKcentral
Cardiac Science Powerheart G5 Green Rescue Ready indicator Red indicator, meaning either a self-test in progress or a detected error Beeps every 20 seconds while the fault persists Daily check of battery, pads and electronics. Weekly partial charge. Monthly full energy charge
Defibtech Lifeline (DDU-2000 series) Blinking green when powered off, solid green when on Blinking or solid red, or no light at all Two beeps periodically while the indicator flashes red Power-on, daily, weekly, monthly and quarterly, plus battery insertion
HeartSine samaritan PAD Flashing green every 5 to 10 seconds Flashing red, or continuous beeping Continuous beeping. Three rapid beeps can mean the unit is outside its temperature range Weekly, at midnight GMT every Sunday

The red that is not a fault

Three situations produce a red indicator on a device that is working correctly. Pulling a unit out of service for any of them wastes time and, worse, trains staff to ignore the indicator.

HeartSine samaritan PAD during its weekly test. The samaritan PAD runs its self-test at midnight GMT every Sunday. The status light blinks red while the test runs and returns to green when it passes. A red flash seen at the wrong moment is not a failure.

Cardiac Science Powerheart during a self-test. On the Powerheart G5 the Rescue Ready indicator turns red both when a self-test is in progress and when an error has been found. The way to tell them apart is the sound: a genuine fault beeps every 20 seconds and persists, a test in progress does not.

Immediately after a battery change. Most units run an insertion self-test as soon as a new battery goes in. Give that test time to complete before reading the indicator.

The rule that resolves all three: a transient red clears on its own within the self-test window. A fault red does not, and on most brands it is paired with a repeating audible alert. If it is still red after the test cycle, treat it as a failure.

The brands where a fault shows as no light at all

This is the failure mode most monthly checks miss, because an unlit indicator reads as “nothing wrong” to anyone walking past.

On the Philips HeartStart OnSite and FRx, ready is a blinking green light and a problem is the light being off. There is no red state. If the light is off and the unit is chirping with the i-button flashing, the cause is a self-test error, a pads problem, a low battery or, on the FRx, an Infant/Child Key left in the device. If the light is off with no chirping and no flashing i-button at all, the likely cause is no battery installed, a fully depleted battery, or a device that needs repair.

On the Stryker LIFEPAK CR2, ready is a readiness indicator that flashes once every 6 seconds. A fault is the indicator simply not flashing, accompanied by an alert tone every 15 minutes. That tone can be switched off in setup, which means a CR2 in a quiet corridor with audio disabled can sit unready and silent indefinitely.

On the Defibtech Lifeline, no light at all is a fault state in its own right and usually means a depleted battery pack.

If your inspection routine says “check the light is green”, it will pass a Philips unit that is completely dead. The check has to be “confirm the indicator is in the ready state for this model”, which is a different instruction.

The ZOLL AED Plus red X

The AED Plus shows its result in a window on the handle: a green check mark means it passed its last self-test and is ready, a red X means it failed and is not ready for use. ZOLL lists the conditions that produce the red X as low or depleted battery capacity, electrode connection problems or missing electrodes, failed ECG circuitry, failed defibrillator charging or discharging circuits, microprocessor or software integrity failures, failed CPR monitoring circuitry, and failed audio circuitry.

Two of those, the battery and the electrodes, are things you can resolve on site. The rest are internal and mean the unit goes back to the manufacturer. The practical sequence is to rule out the two fixable causes first, then stop.

The AED Plus detail most programs get wrong is the schedule. Its self-test runs once every 7 days by default, not daily, though the interval is configurable from 1 to 7 days and software 5.32 and above adds a monthly test. A fault can therefore exist for up to a week before the indicator reflects it, which is the single strongest argument for a physical monthly inspection rather than relying on the window alone.

What to check, in order

1
Confirm the state is real

Check against the table above for your exact model. Wait out the self-test window if the unit has just had a battery fitted or it is a HeartSine on a Sunday.
2
Check the battery

The most common single cause. Confirm the install-by date and that the pack is fully seated. Guidance on this is in the guide to how to check an AED battery.
3
Check the pads

Confirm the cartridge or connector is fully engaged, the pads are in date and the packaging is unbroken. Expired or disconnected pads trigger a fault on most brands.
4
Check for accessories left in place

On the Philips FRx an Infant/Child Key left in the unit is a documented cause of the ready light going off.
5
Run or wait for a fresh self-test

Confirm the indicator returns to the ready state. If it does not, the fault is internal and the unit needs manufacturer service.

Can you use an AED showing a red indicator?

In a cardiac arrest with no alternative device, you use what you have. An AED that has flagged a fault may still deliver a shock, and the alternative is no defibrillation at all. Call emergency services, start CPR and attach the device.

That is an emergency judgement, not a compliance position. A unit showing a fault indicator is not a compliant AED, and in most states Good Samaritan immunity is conditional on the device being maintained as the manufacturer specifies. Leaving a flagged device on the wall is the exposure, not using it in extremis.

Why indicator checks alone fail as a compliance system

The indicator answers one question: did the last self-test pass. It does not tell you that the pads expire in six weeks, that the battery install-by date falls next quarter, or that the two responders trained on that floor both let their certification lapse. None of those will turn the light red until it is already too late to plan around them.

A self-test is a device check. A compliance record is a programme check, and it has to be built from dated inspections, tracked expiry dates for adult and pediatric pads, battery install-by dates, and a current responder list. That is the gap a monthly inspection closes.

Free download

AED Monthly Inspection Checklist

A printable one-page checklist covering status indicator, battery, pads, device condition and monthly sign-off, formatted the way inspectors expect to see it during an audit.

Get the checklist

FAQ

Does a red light always mean the AED is broken?No. On a HeartSine samaritan PAD the status light blinks red while the weekly Sunday self-test runs and returns to green when it passes. On a Cardiac Science Powerheart the indicator is also red while a self-test is in progress. A genuine fault persists past the test window and is normally paired with a repeating audible alert.
My Philips HeartStart has no light at all. Is that better or worse than red?Worse, because it is easy to miss. Philips HeartStart units have no red state. Ready is a blinking green light, and a fault is the light being off. No light with no chirping usually means no battery, a fully depleted battery, or a device needing repair.
Can the defibrillator be used if a red X is in the window?In a genuine emergency with no other device available, yes, use it: call emergency services, start CPR and attach the pads. A red X means the unit failed its self-test and is not considered ready for use, so it should be removed from service and repaired afterwards, and it should never be left on the wall in that state.
What does a red X on a ZOLL AED Plus mean?It means the unit failed its last self-test. ZOLL lists the causes as low or depleted battery, electrode connection problems or missing electrodes, failed ECG circuitry, failed charging or discharging circuits, microprocessor or software integrity failures, failed CPR monitoring circuitry, or failed audio circuitry. Battery and electrodes are fixable on site; the rest require service.
How quickly does the indicator react to a problem?It reacts at the next scheduled self-test, not instantly. The ZOLL AED Plus runs its test every 7 days by default, so a fault can exist for up to a week before the window changes. Philips and several others test daily. This lag is why a physical monthly inspection is still necessary.
The AED is beeping but the light looks normal. What now?Trust the sound. Several brands pair the alert tone with a state that is easy to misread, and on the LIFEPAK CR2 the tone can be disabled in setup entirely. Check the indicator against your model’s documented ready state rather than assuming any lit indicator means ready.

Start tracking AED self-test readiness across every device

Pricing starts at $114 per year for a single AED, set by how many devices you manage. Unlimited users and locations included.

✓ Unlimited locations
✓ 60 & 30-day expiration alerts
✓ Audit-ready inspection logs
Kyle Nowack, Founder of AED Log

About the Author

Kyle Nowack

Founder, AED Log

Kyle Nowack founded AED Log in 2019 to help organizations simplify AED program management. His work focuses on software that tracks AED devices, pad and battery expiration dates, inspection records, and user certifications, helping schools, fitness centers, businesses, and public facilities stay on top of AED readiness.

Through the AED Log blog, Kyle shares practical guidance on device tracking, maintenance planning, and program documentation to help organizations manage their AED responsibilities with greater confidence.

Connect with Kyle on LinkedIn   |   Learn more about AED Log →

Join to newsletter.

Get a personal consultation.

Call us today at (855) 248-2464